Rotronic HygroCal100 Advanced humidity calibrator
The Rotronic HygroCal100 Advanced (HC100A) is a fully integrated, standalone instrument for validating humidity sensors from 5% to 95% RH. This makes it suitable for a wide range of laboratory and process applications, including those in the pharmaceuticals, food and semiconductor sectors.
A re-engineered sealed internal test chamber is designed to provide both stability and uniformity across the chamber, consistent to within ±0.5% RH. With separate saturation and desiccation reservoirs, changing measurement conditions simply requires ambient air to be passed through either reservoir.
Its internal reference, Rotronic’s HC2A-S probe, provides humidity and temperature measurements within 0.8% RH and 0.1°C, with low annual drift. The probe’s onboard memory means the probes can be easily exchanged. Rotronic also offers additional reference device choices for sophisticated users. Users can choose from handheld humidity instruments to precision chilled mirrors.
A colour touchscreen interface enables quick and simple configuration and operation of the HC100A. Once set, the calibration procedures will then run automatically. All measurement and reference data can be stored within the internal memory or downloaded to a PC.
Phone: 03 9017 8225