Rotronic HygroCal100 Advanced humidity calibrator
The Rotronic HygroCal100 Advanced (HC100A) is a fully integrated, standalone instrument for validating humidity sensors from 5% to 95% RH. This makes it suitable for a wide range of laboratory and process applications, including those in the pharmaceuticals, food and semiconductor sectors.
A re-engineered sealed internal test chamber is designed to provide both stability and uniformity across the chamber, consistent to within ±0.5% RH. With separate saturation and desiccation reservoirs, changing measurement conditions simply requires ambient air to be passed through either reservoir.
Its internal reference, Rotronic’s HC2A-S probe, provides humidity and temperature measurements within 0.8% RH and 0.1°C, with low annual drift. The probe’s onboard memory means the probes can be easily exchanged. Rotronic also offers additional reference device choices for sophisticated users. Users can choose from handheld humidity instruments to precision chilled mirrors.
A colour touchscreen interface enables quick and simple configuration and operation of the HC100A. Once set, the calibration procedures will then run automatically. All measurement and reference data can be stored within the internal memory or downloaded to a PC.
Phone: 03 9017 8225
Portable Memosens pH, conductivity and oxygen analyser
The Knick Portavo 904X is an IECEX-approved portable pH, conductivity and oxygen meter for Zone...
Roxtec CSP multi-cable cabinet seal platform
Roxtec CSP multi-cable transits are designed to be area-efficient, adaptable to cables of...
FEAS SSE6030 DC supercapacitor UPS
FEAS GmbH is expanding its SSE30 series with the addition of the 60 V SSE6030...




