Condumax II hydrocarbon and water dew-point analyser increases confidence in gas quality

AMS Instrumentation & Calibration Pty Ltd
Sunday, 21 April, 2013


Michell Instruments enabled a natural gas storage operator to improve the hydrocarbon and water dew-point measurements at its natural gas storage site in Europe.

The storage operator had lost confidence in its current water and hydrocarbon dew-point analysers: the hydrocarbon dew-point readings were erratic and the water dew-point readings were unresponsive. Fast response is critically important to natural gas storage operators providing emergency gas supply to satisfy peak demand on transmission networks - particularly during colder winter months.

The storage operator rented a Condumax II hydrocarbon and water dew-point analyser at short notice to give readings of the natural gas in parallel with the operator’s current analysers. This allowed them to check the accuracy and responsiveness of their equipment against a different analyser at minimum capital expense.

During the test period, the Condumax II provided consistent HCDP readings, and showed better levels of response at lower dew points than the previously installed analysers. The Condumax’s water dew-point readings taken during the test period were significantly faster to changes in the process than the analysers under review.

At the end of the test, the company was satisfied that its concerns about the reliability of the readings of the current analysers were justified. As the engineer explained: “The [previous analyser] drifts to higher HC dew-point levels over time in direct comparison to the consistent and repeatable Condumax HC dew-point readings over several days. This drift behaviour and direction is a problem, because it brings the reading nearer to the agreed contract value for HC dew point with Wingas Transport (Gascade) ...  in [some] cases they were close to breaking the contract value.” Because the customer needed to have complete confidence in the levels of hydrocarbon and water dew point in the process, they chose to install the Condumax II permanently in the process.

Related Articles

Five common mistakes in industrial temperature monitoring

In industrial production, effective temperature and humidity monitoring is more than just...

Video-based water detection reduces contamination in gas pipeline

How a US midstream company improved gas quality and reduced operating costs.

Aquamonix integrates flow monitoring for major gold producer

Instrumentation company Aquamonix has installed a flow monitoring solution for a major...


  • All content Copyright © 2026 Westwick-Farrow Pty Ltd